MINIFLASH TOUCH is a uniquely designed flash point tester for the determination of flashpoints of liquids and solids, using the Grabner flash point detection method of measuring the instantaneous pressure increase inside the continuously closed chamber, triggered by an energy-controlled electric arc. Heating the test chamber from the top avoids condensation of high volatile compounds, significantly improving the test results.
Its continuously closed cup design and small sample volume of only 1-2 mL offers maximum safety, avoids offensive fumes in the testing area and reduces sample waste. The fast thermoelectric cooling at the end of a test and the fast and easy handling of the instrument save expensive labor time.
Furthermore, test results are equivalent to other well known standards such as Abel and TAG, and thanks to its unique design MINIFLASH can be used for more difficult applications such as flash-point testing of contaminated samples like:
The new MINIFLASH TOUCH analyzer also includes improved methods for simulation of EN ISO 3679 and EN ISO 3680.
Features
Technical Data |
|
MINIFLASH FLP TOUCH |
0 to 200°C (32 to 390°F) |
MINIFLASH FLPH TOUCH |
0 to 400°C (´32 to 750°F) |
Power Requirements: |
100/110/230 V AC 50/60 Hz, 150 W |
Field Application: |
12 V/8A DC (vehicle battery) |
W x H x D: |
253 x 368 x 277 mm (10" x 14.5" x 10.9") |
Weight: |
12 kg (26 lb) |
Precision Data |
MINIFLASH |
MINIFLASH |
Pensky-Martens |
Repeatability |
1.9°C (3.4°F) |
3.8°C (6.8°F) |
4.1°C (7.4°F) |
Reproducibility |
3.1°C (5.6°F) |
5.4°C (9.7°F) |
6.9°C (12.4°F) |